Introduction
The instrument series HD34⦠is made up of 4 bench top instruments for electrochemical measures: pH, conductivity,
dissolved oxygen, and temperature.
The displayed data can be stored ( datalogger) and can be transferred to PC or serial printer thanks to the multistandard serial ports RS232C and USB2.0 and software DeltaLog9 ( Vers.2.0 and subsequent ones) . The storing and printing parameters can be set from menu.
The HD3405.2 measures pH, redox potential ( ORP) in mV.
It measures temperature with Pt100 or Pt1000 immersion,
penetration or contact probes. The pH electrode calibration can be carried out on one, two or three points and the calibration sequence can be chosen from a list of 13 buffers.
The HD3406.2 measures conductivity, liquid resistivity in liquids, total dissolved solids ( TDS) and salinity using combined 4-ring and 2-ring conductivity/ temperature probes. Temperature is measured by Pt100 or Pt1000
immersion, penetration or contact probes.
The probe calibration can be performed automatically in one or more of the 147¼ S, 1413¼ S, 12880¼ S or 111800¼ S/ cm conductivity calibration solutions.
The HD3409.2 measures the concentration ( in mg/ l) of dissolved Oxygen in liquids, the saturation index ( in% ) and the temperature using SICRAM combined probes of polarographic type with two or three electrodes and integrated temperature sensor. Temperature is measured by Pt100-SICRAM or direct 4 wire-immersion, penetration
or contact probes.
Thanks to an internal pressure sensor, the instruments automatically compensate for barometric pressure. The
instrument anticipates automatic compensation of the Oxygen probe membrane permeability and of the salinity of the liquid being examined. The dissolved Oxygen probeâ s quick calibration function guarantees timely correctness of the performed measurements.
The HD3456.2 measures pH, mV, redox potential ( ORP) , conductivity, resistivity in liquids, total dissolved solids ( TDS) , and salinity using combined 4-ring and 2-ring conductivity/ temperature probes. Temperature is
measured by Pt100 or Pt1000 immersion, penetration or contact probes.
The pH electrode calibration, as well as manual, can be carried out on one, two or three points and the calibration
sequence can be chosen from a list of 13 buffers.
The probe calibration can be performed automatically in one or more of the 147¼ S, 1413¼ S, 12880¼ S or 111800¼ S/ cm conductivity calibration solutions.
The display shows continually the temperature in ° C or ° F and one selectable parameter according to the connected
probe type, i.e. in case of conductivity probe it is possible to select between à or © or TDS or g/ l.
Other common function of this instrument series include: Max, Min and Avg function, the Auto-HOLD function, the
automatic turning off which can also be disabled.
The instruments have IP66 protection degree.
Technical characteristics of the instrument series HD34â¦
Common technical data
Instrument
Dimensions ( Length x Width x Height) 220x120x55mm
Weight 460g ( complete with batteries)
Materials ABS, rubber
Display 2x4½ characters plus symbols
visible area: 52x42mm
Operating conditions
Working temperature -5 ⦠50° C
Stocking temperature -25 ⦠65° C
Working relative humidity 0 ⦠90% RH without condensation
Protection degree IP66
Power
Batteries 3 batteries 1.5V type AA
Autonomy ( only batteries) 100 hours with 1800mAh alkaline batteries
Mains ( cod. SWD10) Output mains adapter 100-240Vac/ 12vdc-1A
Security of memorized data
Unlimited
Selectable storage interval
1s, 5s, 10s, 15s, 30s, 1min, 2min, 5min, 10min, 15min, 20min, 30min and 1hour
Time
Date and hour Schedule in real time
Accuracy 1min/ month max departure
Serial interface RS232C
Type RS232C electrically isolated
Baud rate Can be set from 1200 to 38400 baud
Data bit 8
Parity None
Stop bit 1
Flow Control Xon/ Xoff
Serial cable length Max 15m
Selectable print interval immediate or 1s, 5s, 10s, 15s, 30s, 1min, 2min, 5min, 10min, 15min, 20min, 30min and 1ora
USB Interface
Type 1.1 - 2.0 electrically isolated
Common connections to all models
Serial interface and USB 8-pole MiniDin connector
Mains adapter ( cod. SWD10) 2-pole connector ( positive at centre) 12Vdc/ 1A
EMC Standard regulations
Security EN61000-4-2, EN61010-1 level 3
Electrostatic discharge EN61000-4-2 level 3
Electric fast transients EN61000-4-4 level 3, EN61000-4-5 level 3
Voltage variations EN61000-4-11
Electromagnetic interference
susceptibility IEC1000-4-3
Electromagnetic interference
emission EN55020 class B
Technical characteristics HD3456.2 pH, mV, à , © , TDS, Sal, ° C/ ° F measurement
Measured values
pH, mV, à , © , TDS, salinity, ° C, ° F
Storage of measured values
Tipo 2000 pages of 10 samples each
Quantity 20, 000 terns of measures made up of [ pH or
mV] , [ à or © or TDS or salinity] and temperature.
Connections
pH/ mV input Female BNC connector
Conductivity input 8-pole male DIN45326 connector
Input for temperature probes with 8-pole male DIN45326 connector
TP47 module
Measurement of pH by Instrument
Measurement range -2.000⦠+ 19.999pH
Resolution 0.01 o 0.001pH selectable from menu
Accuracy ± 0.001pH ± 1digit
Input impedance > 1012©
Calibration error @ 25° C | Offset| > 20mV
Slope > 63mV/ pH or Slope 106.5% or Sensitivity < 85%
Automatic / manual -50⦠+ 150° C
temperature compensation
Measurement of mV by Instrument
Measurement range -1999.9⦠+ 1999.9mV
Resolution 0.1mV
Accuracy ± 0.1mV ± 1digit
Drift after 1 year 0.5mV/ year
Standard solutions automatically detected ( @ 25° C)
1.679pH - 2.000pH - 4.000pH - 4.008pH - 4.010pH - 6.860pH - 6.865pH - 7.000pH - 7.413pH - 7.648pH - 9.180pH - 9.210pH - 10.010pH
Measurement of conductivity by Instrument
Measurement range ( Kcell= 0.01) / Res. 0.000⦠1.999¼ S/ cm / 0.001¼ S/ cm
Measurement range ( Kcell= 0.1) / Res. 0.00⦠19.99¼ S/ cm / 0.01¼ S/ cm
Measurement range ( Kcell= 1) / Res. 0.0⦠199.9¼ S/ cm / 0.1¼ S/ cm
200⦠1999¼ S/ cm / 1¼ S/ cm
2.00⦠19.99mS/ cm / 0.01mS/ cm
20.0⦠199.9mS/ cm / 0.1mS/ cm
Measurement range ( Kcell= 10) / Res. 200⦠1999mS/ cm / 1mS/ cm
Accuracy ( conductivity) ± 0.5% ± 1digit
Measurement of resistivity by Instrument
Measurement range ( Kcell= 0.01) / Res. Up to 1G© Ã
cm / ( * )
Measurement range ( Kcell= 0.1) / Res. Up to 100M© Ã
cm / ( * )
Measurement range ( Kcell= 1) / Res. 5.0⦠199.9© Ã
cm / 0.1© Ã
cm
200⦠999© Ã
cm / 1© Ã
cm
1.00k⦠19.99k© Ã
cm / 0.01k© Ã
cm
20.0k⦠99.9k© Ã
cm / 0.1k© Ã
cm
100k⦠999k© Ã
cm / 1k© Ã
cm
1⦠10M© Ã
cm / 1M© Ã
cm
Measurement range ( Kcell= 10) / Res. 0.5⦠5.0© Ã
cm / 0.1© Ã
cm
Accuracy ( resistivity) ± 0.5% ± 1digit
Measurement of total dissolved solids ( with coeffi cient à / TDS= 0.5)
Measurement range ( Kcell= 0.01) / Res. 0.00⦠1.999mg/ l / 0.005mg/ l
Measurement range ( Kcell= 0.1) / Res. 0.00⦠19.99mg/ l / 0.05mg/ l
Measurement range ( Kcell= 1) / Res. 0.0⦠199.9 mg/ l / 0.5 mg/ l
200⦠1999 mg/ l / 1 mg/ l
2.00⦠19.99 g/ l / 0.01 g/ l
20.0⦠99.9 g/ l / 0.1 g/ l
Measurement range ( Kcell= 10) / Res. 100⦠999 g/ l / 1 g/ l
Accuracy ( total dissolved solids) ± 0.5% ± 1digit
Measurement of salinity
Measurement range / Resolution 0.000⦠1.999g/ l / 1mg/ l
2.00⦠19.99g/ l / 10mg/ l
20.0⦠199.9g/ l / 0.1g/ l
Accuracy ( salinity) ± 0.5% ± 1digit
Automatic/ manual temperature compensation
0⦠100° C with ± T that can be selected from 0.00 to 4.00% / ° C
Reference temperature 20° C o 25° C selectable from menu
à / TDS conversion factor 0.4⦠0.8
Cell constant K ( cm-1) 0.01 - 0.1 - 0.7 - 1.0 - 10.0
Standard solutions automatically detected ( @ 25° C)
147¼ S/ cm
1413¼ S/ cm
12880¼ S/ cm
111800¼ S/ cm
Measurement of temperature by Instrument
Pt100 measurement range -50⦠+ 200° C
Pt1000 measurement range -50⦠+ 200° C
Resolution 0.1° C
Accuracy ± 0.25° C
Drift after 1 year 0.1° C/ year